Graphtec America GL900 - Midi Logger

 Midi Logger

Graphtec America

Rated 0 / 5 (0 reviews)

Category : Data Acquisition

by quote only add to cart

Documents:
Datasheet User's Manual

Overview
Features
Specifications
Calibration

Easy-to-use upright high-speed isolated 8-channel multifunction logger

An easy-to-use upright device enabling isolated 8-channel multifunction input, the GL900 is capable of performing high-speed simultaneous measurements of voltage, temperature, and various other phenomena.

  +/-20 mV to +/-500 V
Thermocouples:
K, J, E, T, R, S, B, N, W
0 to 100%
(the B-530 option is required)
  4 channels
Count, Inst., Rpm
   4 channels
 ‡ Select either Pulse or Logic

*) The input signal needs to be not exceeding the specified voltage at any time. The component in the amplifire of the unit may have the damage by over voltage. Please check the details of over voltage in the specification sheet (Maximum input voltage or Withstand voltage seccion) or the caution label attached on the unit.

Data can be captured to PC-friendly USB memory sticks

Long-term data can be captured directly to built-in 256-MB flash memory or to an external USB memory stick at sampling intervals of from 1 ms to 1 min. For high-speed sampling at intervals faster than 1 ms, up to one million data points can be captured to internal RAM.

Example of 8-channel analog measurement

 

Capture destination 10μs 100μs 500μs 1ms 10ms 100ms 1s
Internal RAM
(up to one million points)
10 seconds Approx. 1 min. and 40 sec. Approx. 8 min. and 20 sec. Approx. 16 min. and 40 sec. Approx. 2 hrs. and 40 sec. Approx. 1 day and 3 hrs. Approx. 11 days and 13 hrs.
Internal flash memory (256 MB) × × × Approx. 1 hour. Approx. 11 hrs. Approx. 4 days Approx. 46 days
External USB memory stick (512 MB) × × × Approx. 22 hrs. Approx. 22 hrs. Approx. 9 days Approx. 93 days
 

The USB memory stick must be a standard model (without fingerprint recognition or other proprietary features).

Can be used as an X-Y recorder

The GL900 reproduces analog X-Y recorder movements and provides the illusion of pen up/pen down movements. It can be operated like an analog X-Y recorder and can also be used as a 4-pen X-Y recorder. The digital data format facilitates post-measurement confirmation of data values and report creation.

High-precision temperature measurement even during high-speed sampling

Lets users perform high-precision temperature measurements even during high-speed sampling – ideal for performing combined voltage and temperature measurements.

Comprehensive built-in trigger and timer functions

Using a combination of trigger and timer functions eliminates superfluous data and enables capture of only the required data.

Setting example 1

To perform measurement over a four-day period starting January 10

Timer setting Date and time Start setting January 10 00 hours 00 minutes
Stop setting January 14 23 hours 59 minutes
Trigger setting Start trigger Off
Stop trigger Off

Setting example 2

To perform measurements of abnormal signals during device operations

 

Timer setting Daily cycle Start setting 09 hours 00 minutes
Stop setting 17 hours 00 minutes
Trigger setting Start trigger Level CH 1 (3 V Rising)
Stop trigger Level CH 1 (2 V Falling)
Repeat On

Setting example 3

To perform measurements every 20 minutes

 

Timer setting Hourly cycle Start setting 00 minutes 00 seconds
Stop setting 20 minutes 00 seconds
Trigger setting Start trigger Off
Stop trigger Off

Setting example 4

To perform measurements for a period of one hour, every four hours, daily

 

Trigger setting Start trigger Off
Stop trigger Scheduled time (one hour)
Repeat On (Repeat interval: 4 hours)

 

Timer settings Timer mode Off, Date and time, Daily cycle, Hourly cycle
Trigger settings Start source setting Off, Level value, External input
Stop source setting Off, Level value, External input, Scheduled time
Pre-trigger 0-100%
Repeat capture On, Off and Repeat interval

High-voltage measurement capability

The wide 500 V range enables 100 to 240 VAC power supply voltage waveform measurements. Using logic input and a clamp meter simultaneously allows measurement of a device’s power supply voltage and current concurrently with sequential control of various points.

Built-in, large-format 5.7-inch color LCD for easy-to-read waveforms

The bright, easy-to-read large-format 5.7-inch color TFT LCD provides vivid, easy-to-read waveform displays. Cursor keys enable fast, easy control and setup. The waveform display can be scrolled at high-speed – 10 ms/DIV.

Free Running display for waveform-checking without the need for data capture

The Free Running display lets users check input signal waveforms even before measurements begin. Since waveforms are displayed on each setup screen, users can make settings while viewing the waveforms.

Easy PC measurement via USB; remote monitoring via Ethernet web server and FTP functions

The USB and Ethernet connections enable transfer of captured data to your PC and setup and control of the GL900 from a PC, even without the PC software provided standard with the GL900.

Web server/FTP server functions

Waveform display and GL900 setup operations can be performed via a web browser (e.g., Internet Explorer). In addition, data files captured to the GL900’s internal memory or to a USB memory stick can be transferred or deleted from the PC.

USB drive mode

When your GL900 is connected to your PC via the USB interface, the GL900 can be operated in USB mode to enable fast, easy data transfers from internal memory to the PC.

NTP client function

Simply connect the GL900 to an NTP server via an Ethernet connection to synchronize GL900 time with NTP server time at periodic intervals

Dedicated software for real-time data capture

Three measurement screens are provided to allow selection of the screen that best suits measurement needs. The Replay screen provides a Zoom screen feature to enable enlarged display of specific sections of long-term measurement data.

Simple operations for anyone

Easy-to-use software using icon keys for intuitive operations

Convenient functions

Various convenient data-processing functions are built in.

  • Direct to Excel function
    This function enables measurement data to be written directly to an Excel file.
     
  • Search function
    This function enables searching for specific values in the captured data.
     
  • CSV batch conversion function
    This function enables batch conversion of multiple captured files to CSV file format.
     
  • Thumbnail function
    This function enables display of captured data files as thumbnails.

 

  • Multifunction input on eight isolated channels
  • High-speed simultaneous sampling on eight channels, 16-bit resolution
  • Equipped with a large-format 5.7 inch color LCD for easy-to-read waveform display
  • Data can also be saved to PC-friendly USB memory stilcks

Do you have a Midi Logger that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your Midi Logger calibrated today! Click Here!

 

There are currently no reviews for this product.

 

Review the Graphtec America GL900 - Midi Logger

 Midi Logger

Name :

Email :

Title :

Review: